General
The AZ 8000 is a Trace Nitrogen Analyser. Its unique design brings reliability and accuracy needed for such measurement. Its own Plasma Emission Detector System extends the lifetime of the cell. This Duty Cycle Controlled System has the property to decrease contamination and coating inside the plasma cell giving a more efficient, reliable and accurate analyser.
Applications
Discover which analyser(s) you need for your analytical requests. Check the application here !
Some Global References
Company | Type of analyser | Model | Impurities |
---|---|---|---|
Air Products (France) | Helium | AZ 8000 | N2 |
Air Liquide (France) | Argon | AZ 8000 | N2 |
Linde (China) | Argon | AZ 8000 | N2 |
Air Liquide (Egypt) | Argon | AZ 8000 | N2 |
Technical specifications
Detector type | Plasma Emission Detector design based on a Duty Cycle Controlled System |
---|---|
Three ranges | |
0 – 1 ppm 0 – 10 ppm 0 – 100 ppm | Display resolution to 0.01 ppm Display resolution to 0.1 ppm Display resolution to 1 ppm Other range possible up to 10000 ppm |
Repeatability | |
Standard features | Manual or autoranging (user selectable) |
Microprocessor controlled | |
5.6” TFT intelligent LCD module with Touch Screen |
|
Self diagnosis system with auto-resolve alarm | |
LAN/Web control | |
4-20 mA isolated output | |
Alarm Historic | |
Safe calibration procedure to avoid any bad calibration | |
Digital ouputs for remote monitoring: (all dry relay contacts) - System status (1 output) - Range in use (3 output) - Calibration in use (1 output) |
|
Options | Internal sampling system for zero, span and sample with remote capabilities |
Serial port: RS-232/422/485/ Pro bus | |
2 alarm outputs (user programmable set point) | |
Gas connections | Sample: 1/8’’ compressing ttings |
Vent | 1/8" compressing fitting |
Calibration gas | Zero: PUR8000 puri ed gas (Getter) |
Span | 8.0 to 9.5 ppm N2 in Argon |
Sample flow requirements | 15 to 200 ml/min |
Flow accuracy | 0 to 200 ml/min +- 1% full scale |
Max operating pressure | 7 bar |
Min. operating pressure | 0.3 bar optional 0.1 bar |
Operating temperature | 10 °C to 45 °C |
Supply | 115 VAC, 50 - 60 Hz or 220 VAC, 50 - 60 Hz |
Power consumption | Maximum 40 watts |
Drift | |
Weight | 13 kg |
Features
Unique Plasma Emission Detector
Large scale measurement
4-20 mA output (standard)
Free maintenance
LAN/Web control
Fast purging time
Low sample consumption
3U cabinet
Find a distributor
Check your closest distributor with our new tool. Click on distributors and it will open a new page.