DID Gas Chromatograph

DID Gas Chromatograph - Helium

Analysis of H2, Ne, O2+Ar, N2, Kr, CH4, CO, CO2 and Xe in Helium, Argon, Hydrogen, Nitrogen, Oxygen, Carbon Dioxide, Krypton and Xenon (other gases possible on request)

General

The DID Gas Chromatograph has been designed to make use of the variations in electrical conductibility found in highly ionised gas. The ionisation is created by a single high frequency discharge within the detector (HF Plasma), causing a high energy photon emission (24.5 eV). This emission is capable of ionising all the gases within the cell with the exception of the Helium.

Principle

One of the biggest advantage of our DID Gas Chromatograph is that a photo-ionised discharge is achieved without any radioactive emissions and without the need to apply excessive voltage being applied to the cell.

Another important characteristic of the DID Gas Chromatograph is that the detector operates using just one gas. This one gas is used as the carrier, the flushing gas and/or the purging gas. This gas can also serve to purge and clean all the valves.

The housing does not require flushing gas: no contamination of the gas circuit is possible before detection as the input connection is located directly on the cell of the DID Gas Chromatograph detector.

The power setting of the high frequency oscillator allows for the adjustment of the detector sensitivity, which can reach 0.1 ppb (parts per billion).

The detection limit and any background noise and directly related to the quality of the carrier gas, so the use of ultra purity (UHP) Helium as the carrier gas is highly recommended.

Some Global References

Company Type of analyser Model Measurement
Linde (China) Helium DID H2, Ne, O2+Ar, N2, CH4
Air Liquide (France) Oxygen DID N2
Linde Worms (Germany) Argon DID / Orthosmart H2, O2, N2, CO, CO2, CH4, NMHC
Linde (Germany) Nitrogen DID+FID H2, O2, Ar, CH4, CO, CO2, NMHC
Air Liquide (France) Krypton DID H2, O2+Ar, N2, Xe

Technical specifications

AccuracyRange and application dependent
Drift
Temperature drift
Operating temperature Around 20°C without wide variations of temperature
Sampling gas He, N2 , H2, O2, Ar, CO2 , Kr, Xe
Sample gas connection1/8’’ Swagelok OR 1/8’’ VCR.
Sample flow rateApproximately 3 to 5 l/h
Sample pressureAtmospheric pressure
Carrier gasHelium
Carrier gas connection1/8’’ Swagelok OR 1/8’’ VCR.
Carrier gas pressure7 bar
Carrier gas flow rate4 l/h
Recommended qualityminimum 6.0
Power supply220 Vac, 50-60 Hz
Power consumption420 VA
4-20 mA outputEight confi gurable outputs
depending on the application.
RJ-45 connectionComputerised system maintenance
Output relays
(SPST 2 amperes / 250 Vac)
1 Analyser Failure Alarm contact
1 Alarm High contact
1 Alarm High High contact

Measurement capabilities

ArgonHydrogenNitrogenOxygenHelium
CH4
CO
H2/
N2/
O2 + Ar//
O2/
Ar/
CO2
Ne
Kr
Xe

Dimensions & chromatogram

(Dimensions can differ regarding the application)

Features

< 10 ppb resolution guaranteed
User-friendly interface
Adjustable alarm and oven settings
Fast response time
Possibility of auto-calibration programming
CE marked
Chromdyne software

Applications